Clinical Orthopaedics and Related Research: February 2013 - Volume 471 - Issue 2 - p 519–526 doi: 10.1007/s11999-012-2559-1 Symposium: Papers Presented at the Annual Meetings of The Hip Society

What Factors Affect Posterior Dislocation Distance in THA?

Nevelos, Jim, PhD1; Johnson, Aaron, MD3; Heffernan, Christopher1; Macintyre, James1; Markel, David, C., MD2; Mont, Michael, A., MD3, a
Hip

Background Dislocation remains common after total hip arthroplasty. Efforts have been made to identify and minimize risk factors. One such factor, jump distance, or the distance the femoral head must travel before dislocating, has been poorly characterized with respect to three-dimensional kinematics.

 

Questions/purposes We therefore determined: (1) the three-dimensional stability of four different component designs; (2) whether the degree of abduction and anteversion affects the stability; (3) whether pelvic inclination angles affected stability; and (4) which combination of these three factors had the greatest stability.

 

Methods We created a positionable three-dimensional model of a THA. Acetabular components were modeled in various abduction and anteversion angles and in two different pelvic inclinations which simulate standing and chair-rising activities.

 

Results The posterior horizontal dislocation distance increased as inclination angle and femoral head size increased. The 48-mm resurfacing typically had lower jump distances and was at risk of posterior edge loading at 30° inclination. The highest jump distance for all positions and activities occurred with the dual-mobility bearing.

 

Conclusion These findings suggest that monoblock cups require extremely accurate positioning for low dislocation risk and that pelvic orientation may increase dislocation risks.

 

Clinical Relevance As a result of the dual-mobility designs having the greatest resistance to dislocation, these cups may be appropriate for patients who are at risk for dislocation in difficult primary situations and in revision hip arthroplasty procedures in which proper component orientation may be less likely to be achieved.