The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 23, Issue: 7, Page: 1068-1070

The Combined Anteversion Technique for Acetabular Component Anteversion

Amuwa, Christopher; Dorr, Lawrence D
Hip

The combined anteversion technique for acetabular component placement of total hip arthroplasty is beneficial because of the surgeons’ limited ability to control the anteversion of a cementless femoral stem. Our data show that the cementless stem anteversion can be 15° different than anticipated. By determining femoral stem anteversion before positioning cup anteversion, the cup anteversion can be adjusted for the stem anteversion. The combined anteversion technique should provide a mean near 35° with a safe zone of 25° to 50°.


Link to article