© 2018 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 36:1998–2006, 2018.

Comparing damage on retrieved total elbow replacement bushings with lab worn specimens subjected to varied loading conditions

Ryan Willing
Elbow

Complication rates following total elbow replacement (TER) with conventional implants are relatively high due to mechanical failure involving the UHMWPE bushings. Unfortunately, there are no standardized pre‐clinical durability testing protocols for assessing the durability of TER components. This study examines the damage observed on retrieved humeral bushings, and then uses in vitro durability testing with two different loading protocols to compare resulting damage. Damage on 25 pairs of retrieved humeral bushings was characterized using micro‐computed tomographic imaging techniques. The damage was compared with that of in vitro test specimens which were subjected to 200 K cycles of either high joint reaction force (high JRF) or high varus moment (high VM) loading. Material removal (mass loss) from bushing components was measured using gravimetric techniques. Thinning was less for retrieved bushings which were still assembled in their humeral component, versus bushings which were loose (0.3 ± 0.3 mm vs. 0.6 ± 0.3 mm, p = 0.02). Comparing in vitro test specimens, thinning due to high VM loading was 0.9 ± 0.3 mm, versus 0.2 ± 0.0 mm for high JRF loading (p = 0.08); however, the actual material removal rates from the humeral bushings were not different between the two protocols (48 ± 5 mm3/Mc vs. 43 ± 2 mm3/Mc, p = 1). Neither loading protocol could produce damage patterns fully representative of the spectrum of damage patterns observed on clinical retrievals. Pre‐clinical testing should employ multiple loading protocols to characterize implant performance under a broader spectrum of usage.


Link to article