The Journal of Arthroplasty, Volume 28, Issue 3, 521 - 525

Cement-in-Cement Revision for Selected Vancouver Type B1 Femoral Periprosthetic Fractures: A Biomechanical Analysis

Brew, Christopher J. et al.
Hip

The aim of this study was to perform a biomechanical analysis of the cement-in-cement (c-in-c) technique for fixation of selected Vancouver Type B1 femoral periprosthetic fractures and to assess the degree of cement interposition at the fracture site. Six embalmed cadaveric femora were implanted with a cemented femoral stem. Vancouver Type B1 fractures were created by applying a combined axial and rotational load to failure. The femora were repaired using the c-in-c technique and reloaded to failure. The mean primary fracture torque was 117 Nm (SD 16.6, range 89–133). The mean revision fracture torque was 50 Nm (SD 16.6, range 29–74), which is above the torque previously observed for activities of daily living. Cement interposition at the fracture site was found to be minimal.


Link to article