The influence of early radiolucent lines appearing on femoral head penetration into HXLPE cemented sockets. HIP International, 28(4), 369–374.

The influence of early radiolucent lines appearing on femoral head penetration into HXLPE cemented sockets

Morita, D., Seki, T., Higuchi, Y., Takegami, Y., Amano, T., & Ishiguro, N. (2018).
Hip

This study investigates differences in femoral head penetration between highly cross-linked polyethylene (HXLPE) cemented sockets both with and without radiolucent lines (RLLs) in the early postoperative phase and at 5 years follow-up.

There were 35 patients (37 hips), mean age of 66.8 years, who underwent total hip arthroplasty (THA) using highly HXLPE cemented sockets. They were divided into 2 groups based on postoperative the early appearance of RLLs. Femoral head penetrations on both anteroposterior- and Lauenstein-view radiographs were evaluated, and the mean polyethylene (PE) wear rate was calculated based on femoral head penetrations between 2 and 5 years.

Femoral head penetrations in the proximal direction were 0.075 mm and 0.150 mm in the RLL and non-RLL groups at 1 year postoperatively (p = 0.019). At 5 years measured penetration was 0.107 mm and 0.125 mm in the RLL and non-RLL groups, respectively (p = 0.320). The mean PE wear rates in anteroposterior-view were 0.008 mm/year and 0.003 mm/year in the RLL and non-RLL groups (p = 0.390) and those in Lauenstein-view were 0.010 mm/year and 0.005 mm/year, respectively (p = 0.239).

In the RLL group, the PE bedding-in was less compared with those in the non-RLL group. Additionally, the mean PE wear rate in the RLL group tended to be higher than that in the non-RLL group. The distribution of stress loading through the cement may differ according to whether early RLLs appear.