The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 25, Issue: 3, Page: 497.e7-497.e10

Double-Concave Deformity of the Polyethylene Tibial Post in Posterior Stabilized Total Knee Arthroplasty

Niki, Yasuo; Matsumoto, Hideo; Yoshimine, Fumihiro; Toyama, Yoshiaki; Suda, Yasunori; Banks, Scott A
Knee
This report describes a unique case of bilateral total knee arthroplasty necessitating revision of the polyethylene insert, which showed prominent marks on the tibial post resulting from repeated seiza-style sitting. The patient presented 7 years postoperatively with knee pain and flexion disturbance due to continuous joint effusion persisting for more than 4 months. Proliferating synovia throughout the joint revealed reactive synovitis to polyethylene particles. The retrieved polyethylene inserts displayed double-concave deformity of the tibial post with burnishing and creep in tibiofemoral articulation. The damage pattern of retrieved polyethylene inserts reflected the data from tibiofemoral contact location obtained using a shape-matching technique in the early postoperative phase. This case provides an example of damage to the polyethylene tibial post caused by a floor-sitting lifestyle and the potential clinical sequelae.