The Journal of Arthroplasty, ISSN: 0883-5403, Vol: 25, Issue: 5, Page: 680-686

Comparison of Steady State Femoral Head Penetration Rates Between Two Highly Cross-Linked Polyethylenes in Total Hip Arthroplasty

Whittaker, John-Paul; Charron, Kory D; McCalden, Richard W; Macdonald, Steven J; Bourne, Robert B
Hip

Given that the manufacture of highly cross-linked polyethylene (HXLPE) is not standardized, the behavior of these materials may vary. Our study compares minimum 5-year steady state femoral head penetration rates using the Martell method, in 2 HXPLEs produced by different manufacturers. Patients received a primary hip arthroplasty using an uncemented acetabular component with an HXLPE liner and a 28-mm femoral head. Forty-seven patients in group A received an HXLPE liner (Reflection XLPE, Smith and Nephew Inc, Memphis, Tenn), and 36 patients in group B received a different HXLPE liner (Longevity, Zimmer Inc, Warsaw, Ind). Average follow-up was 6.42 years in group A and 7.64 years in group B. The steady state head penetration rates were not significantly (P > .05) different between the HXPLE groups over the midterm with 0.026 mm/y and 0.025 mm/y in groups A and B, respectively.


Link to article