International Orthopaedics May 2012, Volume 36, Issue 5, pp 915–920

The Bicontact cementless primary total hip arthroplasty: long-term results

Swamy, G., Pace, A., Quah, C. et al.
Hip

We report on a prospective series of 201 primary uncemented total hip arthroplasties with a Bicontact prosthesis at a mean follow-up of 12.9 years. The mean Harris hip score improved from 41 before surgery to 89 at final follow-up. Two femoral stems, one each for infection and fracture, and 12 cups were revised. The mean number of years to revision post-primary surgery was 8.7 years (six months to 16 years). The cumulative survival of the prosthesis was 95.42% for any cause at ten years and 93.57% at 12 years. Survivorship for aseptic loosening of the femoral stem was 100%. In our series, the Bicontact uncemented total hip arthroplasty stem without hydroxyapatite coating showed excellent survival and the cup survival was comparable to other leading series.