Size of cup affects the anterior capsular distance in total hip arthroplasty, as measured with ultrasound. BMC Musculoskelet Disord 15, 23 (2014).

Size of cup affects the anterior capsular distance in total hip arthroplasty, as measured with ultrasound

Tarasevicius, S., Loiba, V., Stucinskas, J. et al.
Hip

Background

Previously was found that sonography is a reliable method to measure a capsular distance in total hip arthroplasty hips. The aim of our current study was to investigate the relation between the implanted size of the cup and the anterior capsular distance, as measured with ultrasound one year after THA.

Methods

50 osteoarthritis (OA) patients operated on with total hip arthroplasty one year before were included in the study and the anterior capsular distance was measured sonographically. Patients were grouped with respect to cup size. The correlation between the implanted cup size and capsular distance was determined.

Results

The mean capsular distance in the whole group was 1.37 (SD 0.19) cm. The mean capsular distance in the group with small cups was 1.27 (SD 0.13) cm, in large cups it was 1.45 (SD 0.20) cm, p = 0.02. Spearman correlation analysis showed a statistically significant correlation between a greater capsular distance and the larger size of the cup (r = 0.5, p < 0.0001).

Conclusion

The greater capsular distance in successful THA hips is affected by cup size. We propose that this should be considered when evaluating sonography of the anterior capsular distance after THA.