Pseudotumours, cobalt and clinical outcome in small head metal-on-metal versus conventional metal-on-polyethylene total hip arthroplasty. HIP International, 30(1), 56–63.

Pseudotumours, cobalt and clinical outcome in small head metal-on-metal versus conventional metal-on-polyethylene total hip arthroplasty

van der Veen, H. C., Reininga, I. H., Zijlstra, W. P., Boomsma, M. F., Bulstra, S. K., & van Raay, J. J. (2020).
Hip

Metal-on-metal total hip arthroplasty (MoM THA) is associated with the formation of pseudotumours. Studies mainly concern pseudotumour formation in large head MoM THA. We performed a long-term follow-up study, comparing pseudotumour incidence in small head metal-on-metal (SHMoM) THA with conventional metal-on-polyethylene (MoP) THA. Predisposing factors to pseudotumour formation were assessed.

From a previous randomised controlled trial comparing SHMoM (28 mm) cemented THA with conventional MoP cemented THA, patients were screened using a standardised CT protocol for the presence of pseudotumours. Serum cobalt levels and functional outcome were assessed.

56 patients (33 MoP and 23 MoM) were recruited after mean follow-up of 13.4 years (SD 0.5). The incidence of pseudotumours was 1 (5%) in the SHMoM THA cohort and 3 (9%) in the MoP THA cohort. Prosthesis survival was 96% for both SHMoM and MoP THAs. Serum cobalt levels did not exceed acceptable clinical values (<5 µg/L) whereas no differences in cobalt levels were detected at follow-up between both groups. Oxford and Harris Hip Scores were good and did not differ between SHMoM and MoP THA.

This long-term follow-up study shows a low incidence of pseudotumour formation and good functional outcome in cemented head-taper matched SHMoM and MoP THA.


Link to article