Knee Surgery, Sports Traumatology, Arthroscopy April 2019, Volume 27, Issue 4, pp 1241–1250

Patellofemoral design enhancements reduce long-term complications of postero-stabilized total knee arthroplasty

Jan, N., Fontaine, C., Migaud, H. et al.
Knee

Purpose

Few studies investigated whether trochlear and patellar design enhancements improve long-term outcomes of total knee arthroplasty (TKA). This study aimed to compare the long-term survival and complication rates of two consecutive generations of the same TKA system with identical tibiofemoral geometry, but different patellofemoral designs.

 

Methods

The authors retrieved the records of 93 patients (104 knees) operated with the HLS II system and 116 patients (122 knees) operated with HLS Evolution system. Patients were evaluated preoperatively and at a minimum of 10 years noting all complications. Kaplan–Meier (KM) survival was compared for two endpoints: (1) revision of all components and (2) revision of any component.

 

Results

From the HLS II series, the incidence of revision of all components was 6.4%, and of any component was 9.8%. From the HLS Evolution series, the incidence of revision of all components was 4.1%, and of any component was 5.1%. Comparing the survival at equivalent follow-up of 14 years, considering revision of all components, the HLS II had higher survival than the HLS Evolution (98.9% vs 95.9%), while considering revision of any component, the HLS II had lower survival than the HLS Evolution (93.0% vs 94.9%). The differences in survival of the two implants were not significant, neither at equivalent follow-up of 14 years (n.s.), nor at maximum follow-up of each cohort (n.s.). The complication rate was higher for the HLS II series compared to the HLS Evolution (28% vs 12%, p = 0.009), but patellofemoral complications were not more frequent (8% vs 6%, n.s.).

 

Conclusions

Though the differences in survival of the two implants were not significant, conflicting findings are observed due to partial revisions for patellar fractures (5 in the HLS II series and 1 in the HLS Evolution series) which could be related to patellofemoral design enhancements. This study highlights the importance of patello-femoral geometry, which is often overlooked in TKA.

 

Level of evidence

Retrospective comparative study, Level III.