CoxaPro
> Clinical Library > Tervetuloa Clinical Libraryyn > Migration and wear of the Duraloc “1200 series” cup associated with Enduron UHWMPE using the EBRA method and the Imagika software
Migration and wear of the Duraloc “1200 series” cup associated with Enduron UHWMPE using the EBRA method and the Imagika software. HIP International. 2010;20(2):198-203.
Hip
Link to article
Migration and wear of the Duraloc “1200 series” cup associated with Enduron UHWMPE using the EBRA method and the Imagika software
Thirion T, Georis P, Gillet P.Hip
We report a retrospective study on 122 THA using Duraloc “1200 series” cups. All patients received a 28-mm head (107 Al2O3, 15 CoCr). The mean age of the patients at surgery was 56.3 years and the mean follow-up was 137 months.
Mean migration at two years using the EBRA method was 0.18 mm cranially and 0.16 mm medially. Mean polyethylene wear was 0.03 mm/year using the Imagika software and 0.06 mm/year using EBRA. Only one patient presented with abnormal polyethylene wear of 1.7 mm at 90 months with Imagika and 2.4 mm with EBRA. The Duraloc cup seems to be a reliable implant in respect of bone fixation and resistance to wear.
Link to article