The Journal of Arthroplasty, Volume 29, Issue 5, 1021 - 1025

Femoral Pinless Length and Offset Measurements During Computer-Assisted, Minimally Invasive Total Hip Arthroplasty

Renkawitz, Tobias et al.
Hip

We asked whether the intraoperative assessment of leg length (LL) and offset (OS) change would be accurate using a novel pinless femoral reference system during unilateral minimally invasive THA in 50 patients with a mean age of 60 years (48–79). LL and OS change measured at surgery was compared with LL/OS change as measured on magnification-corrected preoperative and postoperative radiographs by two blinded examiners. The radiographic evaluation showed a high inter-rater reliability (r > 0.80 for all assessments). The mean differences (±95% limits of agreement) between navigation and radiographic measurements on the treated side were +0.4 mm (±3.6) for LL and −1.0 mm (±3.9) for OS. Femoral pinless navigation technology represents a feasible assistance in THA.


Link to article