The Journal of Arthroplasty, Volume 28, Issue 10, 1722 - 1727

Factors in High-Flex Posterior Stabilized Fixed-Bearing Total Knee Arthroplasty Affecting In Vivo Kinematics and Anterior Tibial Post Impingement During Gait

Tamaki, Masashi et al.
Knee

One of the most important issues in high-flex posterior stabilized total knee arthroplasty (TKA) is anterior tibial post impingement. We examined 20 knees and 2 TKA designs to evaluate the in vivo kinematics during gait using fluoroscopic image analysis. The AP positions in both stance and swing phases were significantly larger and more anterior in J-curved design TKA compared with Single radius design TKA. Anterior tibial post impingement was observed in five knees implanted with Legacy-flex, and in four of the five knees they were not hyperextended. No impingement was observed with NRG. Less constraint and tibial posterior slope might lead to large AP translations and anterior tibial post impingement. Implant design and operative procedures for high-flexion TKAs affect the in vivo kinematics and the occurrence of anterior tibial post impingement.


Link to article