Factors Affecting Squeaking in Metal on Metal Hip Resurfacings. HIP International. 2014;24(4):340-346.

Factors Affecting Squeaking in Metal on Metal Hip Resurfacings

Imbuldeniya A, Munir S, Chow J, Walter WL, Zicat BA, Walter WK.
Hip

We reviewed 380 hip resurfacings cases between December 1999 and Dec 2012. 11 cases (2.89%) squeaked postoperatively. Mean follow-up was 88.6 months (19-130 months). Mean time to squeak was 11.3 months (3-22 months). Ten (91%) cases were male and nine (81%) cases had a Birmingham hip resurfacing. Cases were matched for age, gender, BMI and implant to three controls. Radiographs were analysed using EBRA (Einzel-Bild-Roentgen-Analysis, University of Innsbruck, Austria) software to evaluate cup orientation. There was no significant difference between the mean inclination angle of the cups (p = 0.26) or the mean anteversion angle (p = 0.29). There was no difference in serum cobalt (p = 0.20) or serum chromium (p = 0.45) levels at latest follow-up. Three of the 11 (27.3%) cases had revision surgery at a mean follow-up of 101 months (72-117 months). Squeaking was not influenced by patient demographic parameters and resolved in all cases that did not undergo revision surgery. Males with a resurfacing head size <50 mm had significantly increased odds of squeaking when compared to controls (odds ratio = 26.6; 95% CI = 1.2-573.3; p<0.05). Based on our findings, we do not recommend that squeaking on its own should be an indication for revision of hip resurfacing components.