Excellent Mid-Term Results of a New Polished Tapered Modular Cemented Stem: A Study of 113 Hip Replacements with Minimum 5-Year Follow-Up. HIP International, 27(5), 455–459.

Excellent Mid-Term Results of a New Polished Tapered Modular Cemented Stem: A Study of 113 Hip Replacements with Minimum 5-Year Follow-Up

Vanbiervliet, J., & Somers, J. F. A. (2017).
Hip

This prospective cohort study reports the medium-term clinical and radiographic results of 113 hips treated with a hybrid total hip replacement (THR) with a new cemented tapered cobalt-chrome (Co-Cr) stem with a titanium (Ti) modular neck (ProfemurXm®).

Between October 2008 and December 2010 we performed 115 consecutive hybrid THR with the ProfemurXm® in 105 patients.

Survivorship of the implant (stem and modular neck) at a mean of 6.5 years (min 5-max 8) was 100% with the endpoint revision for any reason. No implant was at risk for revision or showed signs of loosening. The mean Harris Hip Score was 89/100, mean Oxford Hip Score was 43/48, mean WOMAC was 91/100. No patient had thigh pain, no patient reported squeaking. There were no dislocations in this cohort. No implant showed development of radiolucent lines (RLL), either at the stem-cement or cement-bone interface. No hip showed osteolysis or calcar resorption. The mean femoral subsidence of the stem within the cement mantel was 0.31 mm (range 0-0.6 mm) after 6.5 years. With the use of this modular stem, 93% of hips showed no measurable leg length difference after THR, and leg length could be restored within a 5-mm limit in 99% of hips.

The mid-term results of this new polished stem were excellent, without adverse effects from the use of modularity.


Link to article