Acta Orthopaedica, 87:2, 100-105, DOI: 10.3109/17453674.2015.1103607

Excellent long-term results of the Müller acetabular reinforcement ring in primary total hip arthroplasty

Aurimas Sirka, Martin Clauss, Sarunas Tarasevicius, Hans Wingstrand, Justinas Stucinskas, Otto Robertsson, Peter Emil Ochsner & Thomas Ilchmann
Hip

Background and purpose — The original Müller acetabular reinforcement ring (ARR) shows favorable medium-term results for acetabular reconstruction in total hip arthroplasty, where it is used when the acetabular bone stock is deficient. However, there are no data regarding long-term survival of the device. We therefore investigated long-term survival and analyzed radiological modes of failure.

Patients and methods — Between 1984 and 2002, 321 consecutive primary arthroplasties using an ARR were performed in 291 patients. The mean follow-up time was 11 (0–25) years, and 24 hips were lost to follow-up. For survival analysis, we investigated 321 hips and the end of the follow-up was the date of revision, date of death, or the last patient contact date with implant still in situ. Radiological assessment was performed for 160 hips with a minimum of 10 years of follow-up and with radiographs of sufficient quality. It included evaluation of osteolysis, migration, and loosening.

Results — 12 ARR THAs were revised: 1 isolated ARR revision for aseptic loosening, 4 revisions of the ARR and the stem for aseptic loosening, 6 for infection, and 1 for recurrent dislocation. The cumulative revision rate for all components, for any reason, at 20 years was 15% (95% CI: 10–22), while for the ARR only it was 7% (95% CI: 4–12) for any reason and 3.4% (95% CI: 1–9) for aseptic loosening. 21 (13%) of 160 ARR THAs examined had radiological changes: 7 had osteolysis but were not loose, and 14 were radiologically loose but were not painful and not revised.

Interpretation — Our data suggest that the long-term survival of the ARR is excellent.


Link to article