The Journal of Arthroplasty, Volume 27, Issue 4, 539 - 544

Disappointing Short-Term Results With the DePuy ASR XL Metal-on-Metal Total Hip Arthroplasty

Bernthal, Nicholas M. et al.
Hip

Outcomes of ultralarge-diameter femoral heads used in metal-on-metal (MOM) total hip arthroplasty (THA) are relatively unknown. This study reports on early failures of the ASR XL (Depuy, Warsaw, Ind) and assesses whether a correlation with cup positioning exists. A retrospective review of 70 consecutive MOM THAs with ultralarge-diameter femoral head and monoblock acetabular component was conducted. Minimum follow-up was 24 months. Of 70 THAs, 12 (17.1%) required revision within 3 years for pain (7), loosening (3), and squeaking (2). Three additional THAs noted squeaking, 2 noted grinding, and 3 additional hips had persistent pain. In total, 20 (28.6%) of 70 demonstrated implant dysfunction. Acetabular components for all symptomatic hips were in acceptable range of cup abduction and anteversion. The failures noted with this design do not correlate to cup placement. The high rate of implant dysfunction at early follow-up suggests serious concerns with the concept of MOM THA with an ultralarge-diameter femoral head paired with a monoblock acetabular cup.