The Knee, ISSN: 1873-5800, Vol: 19, Issue: 5, Page: 611-6

Computer guided restoration of joint line and femoral offset in cruciate substituting total knee arthroplasty

Shetty, Gautam M; Mullaji, Arun; Bhayde, Sagar
Knee

Purpose

This prospective study aimed to evaluate radiographically, change in joint line and femoral condylar offset with the optimized gap balancing technique in computer-assisted, primary, cruciate-substituting total knee arthroplasties (TKAs).

Methods

One hundred and twenty-nine consecutive computer-assisted TKAs were evaluated radiographically using pre- and postoperative full-length standing hip-to-ankle, antero-posterior and lateral radiographs to assess change in knee deformity, joint line height and posterior condylar offset.

Results

In 49% of knees, there was a net decrease (mean 2.2 mm, range 0.2–8.4 mm) in joint line height postoperatively whereas 46.5% of knees had a net increase in joint line height (mean 2.5 mm, range 0.2–11.2 mm). In 93% of the knees, joint line was restored to within ± 5 mm of preoperative values. In 53% of knees, there was a net increase (mean 2.9 mm, range 0.2–12 mm) in posterior offset postoperatively whereas 40% of knees had a net decrease in posterior offset (mean 4.2 mm, range 0.6–20 mm). In 82% of knees, the posterior offset was restored within ± 5 mm of preoperative values.

Conclusions

Based on radiographic evaluation in extension and at 30° flexion, the current study clearly demonstrates that joint line and posterior femoral condylar offset can be restored in the majority of computer-assisted, cruciate-substituting TKAs to within 5 mm of their preoperative value. The optimized gap balancing feature of the computer software allows the surgeon to simulate the effect of simultaneously adjusting femoral component size, position and distal femoral resection level on joint line and posterior femoral offset.

Level of Evidence

Level II

Link to article