Acta Orthopaedica, 86:4, 477-484, DOI: 10.3109/17453674.2015.1012975

Comparison of the risk of revision in cementless total hip arthroplasty with ceramic-on-ceramic and metal-on-polyethylene bearings

Claus Varnum, Alma B Pedersen, Per Kjærsgaard-Andersen & Søren Overgaard
Hip

Background and purpose — Ceramic-on-ceramic (CoC) bearings were introduced in total hip arthroplasty (THA) to reduce problems related to polyethylene wear. We compared the 9-year revision risk for cementless CoC THA and for cementless metal-on-polyethylene (MoP) THA.

Patients and methods — In this prospective, population-based study from the Danish Hip Arthroplasty Registry, we identified all the primary cementless THAs that had been performed from 2002 through 2009 (n = 25,656). Of these, 1,773 THAs with CoC bearings and 9,323 THAs with MoP bearings were included in the study. To estimate the relative risk (RR) of revision, we used regression with the pseudo-value approach and treated death as a competing risk.

Results — 444 revisions were identified: 4.0% for CoC THA (71 of 1,773) and 4.0% for MoP THA (373 of 9,323). No statistically significant difference in the risk of revision for any reason was found for CoC and MoP bearings after 9 years of follow-up (adjusted RR = 1.3, 95% CI: 0.72–2.4). Revision rates due to component failure were 0.5% (n = 8) for CoC bearings and 0.1% (n = 6) for MoP bearings (p < 0.001). 6 patients with CoC bearings (0.34%) underwent revision due to ceramic fracture.

Interpretation — When compared to the “standard” MoP bearings, CoC THA had a 33% higher (though not statistically significantly higher) risk of revision for any reason at 9 years.


Link to article