The Journal of Arthroplasty, Volume 32, Issue 8, 2386 - 2389

Carbon Dioxide Exposure Resulting From Hood Protective Equipment Used in Joint Arthroplasty Surgery

Patel, Suhani et al.
Hip Knee

Background

To protect both the surgeon and patient during procedures, hooded protection shields are used during joint arthroplasty procedures. Headache, malaise, and dizziness, consistent with increased carbon dioxide (CO2) exposure, have been anecdotally reported by surgeons using hoods. We hypothesized that increased CO2concentrations were causing reported symptoms.

Methods

Six healthy subjects (4 men) donned hooded protection, fan at the highest setting. Arm cycle ergometry at workloads of 12 and 25 watts (W) simulated workloads encountered during arthroplasty. Inspired O2 and CO2concentrations at the nares were continuously measured at rest, 12 W, and 25 W. At each activity level, the fan was deactivated and the times for CO2 to reach 0.5% and 1.0% were measured.

Results

At rest, inspired CO2 was 0.14% ± 0.04%. Exercise had significant effect on CO2 compared with rest (0.26% ± 0.08% at 12 W, P = .04; 0.31% ± 0.05% at 25 W, P = .003). Inspired CO2 concentration increased rapidly with fan deactivation, with the time for CO2 to increase to 0.5% and 1.0% after fan deactivation being rapid but variable (0.5%, 12 ± 9 seconds; 1%, 26 ± 15 seconds). Time for CO2 to return below 0.5% after fan reactivation was 20 ± 37 seconds.

Conclusion

During simulated joint arthroplasty, CO2 remained within Occupational Safety and Health Administration (OSHA) standards with the fan at the highest setting. With fan deactivation, CO2 concentration rapidly exceeds OSHA standards.