The Knee, ISSN: 1873-5800, Vol: 21, Issue: 2, Page: 406-9

A comparison of alignment using patient specific guides, computer navigation and conventional instrumentation in total knee arthroplasty

MacDessi, Samuel J; Jang, Bob; Harris, Ian A; Wheatley, Emma; Bryant, Carl; Chen, Darren B
Knee

Background

Patient specific guides (PSG) have been introduced as a tool in total knee arthroplasty (TKA) in an attempt to improve limb alignment and reduce operative time compared to other established surgical techniques. The purpose of this study was to compare the post-operative radiographic alignment and operative time in patients who underwent TKA surgery with PSG, conventional instrumentation or computer-assisted navigation surgery using fully cemented components.

Methods

A cohort of 260 patients who underwent TKA surgery using PSG (PSG group, n = 115) was compared to patients who underwent TKA using either conventional instrumentation (CON group, n = 92) or computer-assisted navigation (CAS group, n = 53). Post-operative CT imaging using the Perth CT protocol was used to compare alignment between the three groups.

Results

In the PSG and CAS groups, the post-operative hip–knee angle (HKA) was within 3° of neutral alignment in 91.3% and 90.7% of patients, respectively. This compared to 80.4% of patients in the CON group ( p = 0.02). There were no significant differences with respect to alignment when comparing individual component positioning between the PSG and CAS groups apart from tibial slope (Table 3). Total operative time was found to be significantly reduced in the PSG group (80.2 min) compared to both the CON group (86 min, p = 0.002) and the CAS group (110.2 min, p < 0.0001).

Conclusions

The use of PSG resulted in similar alignment accuracy to CAS and superior alignment to CON with significantly shorter operative times.