The Journal of Arthroplasty, Volume 27, Issue 4, 643 - 647

Scaling Digital Radiographs for Templating in Total Hip Arthroplasty Using Conventional Acetate Templates Independent of Calibration Markers

Brew, Christopher J. et al.
Hip

We describe a scaling method for templating digital radiographs using conventional acetate templates independent of template magnification without the need for a calibration marker. The mean magnification factor for the radiology department was determined (119.8%; range, 117%-123.4%). This fixed magnification factor was used to scale the radiographs by the method described. Thirty-two femoral heads on postoperative total hip arthroplasty radiographs were then measured and compared with the actual size. The mean absolute accuracy was within 0.5% of actual head size (range, 0%-3%) with a mean absolute difference of 0.16 mm (range, 0-1 mm; SD, 0.26 mm). Intraclass correlation coefficient showed excellent reliability for both interobserver and intraobserver measurements with intraclass correlation coefficient scores of 0.993 (95% CI, 0.988-0.996) for interobserver measurements and intraobserver measurements ranging between 0.990 and 0.993 (95% CI, 0.980-0.997).