The Journal of Arthroplasty, Volume 32, Issue 6, 2005 - 2011

Does Maximal External Tibial Component Rotation Influence Tibiofemoral Load Distribution in the Primary Knee Arthroplasty Setting: A Comparison of Neutral vs Maximal Anatomical External Rotatory States

Manning, William A. et al.
Knee

Background

Tibial component rotation at time of knee arthroplasty can influence conformity, load transmission across the polyethylene surface, and perhaps ultimately determined survivorship. Optimal tibial component rotation on the cut surface is reliant on standard per operative manual stressing. This subjective assessment aims to balance constraint and stability of the articulation through a full arc of movement.

Methods

Using a cadaveric model, computer navigation and under defined, previously validated loaded conditions mimicking the in vivo setting, the influence of maximal tibial component external rotation compared with the neutral state was examined for changes in laxity and tibiofemoral continuous load using 3D displacement measurement and an orthosensor continuous load sensor implanted within the polyethylene spacer in a simulated single radius total knee arthroplasty.

Results

No significant difference was found throughout arc of motion (0-115 degrees of flexion) for maximal varus and/or valgus or rotatory laxity between the 2 states. The neutral state achieved equivalence for mediolateral load distribution at each point of flexion. We have found that external rotation of the tibial component increased medial compartment load in comparison with the neutral position. Compared with the neutral state, external rotation consistently effected a marginal, but not significant reduction in lateral load under similar loading conditions. The effects were most pronounced in midflexion.

Conclusion

On the basis of these findings, we would advocate for the midtibial tubercle point to determine tibial component rotation and caution against component external rotation.