Acta Orthopaedica, 85:4, 375-382, DOI: 10.3109/17453674.2014.908345

A comparison of the diagnostic accuracy of MARS MRI and ultrasound of the painful metal-on-metal hip arthroplasty

Imran A Siddiqui, Shiraz A Sabah, Keshthra Satchithananda, Adrian K Lim, Suzie Cro, Johann Henckel, John A Skinner & Alister J Hart
Hip

Background and purpose

 Metal artifact reduction sequence (MARS) MRI and ultrasound scanning (USS) can both be used to detect pseudotumors, abductor muscle atrophy, and tendinous pathology in patients with painful metal-on-metal (MOM) hip arthroplasty. We wanted to determine the diagnostic test characteristics of USS using MARS MRI as a reference for detection of pseudotumors and muscle atrophy.

Patients and methods

 We performed a prospective cohort study to compare MARS MRI and USS findings in 19 consecutive patients with unilateral MOM hips. Protocolized USS was performed by consultant musculoskeletal radiologists who were blinded regarding clinical details. Reports were independently compared with MARS MRI, the imaging gold standard, to calculate predictive values.

Results

 The prevalence of pseudotumors on MARS MRI was 68% (95% CI: 43–87) and on USS it was 53% (CI: 29–76). The sensitivity of USS in detecting pseudotumors was 69% (CI 39–91) and the specificity was 83% (CI: 36–97). The sensitivity of detection of abductor muscle atrophy was 47% (CI: 24–71). In addition, joint effusion was detected in 10 cases by USS and none were seen by MARS MRI.

Interpretation

 We found a poor agreement between USS and MARS MRI. USS was inferior to MARS MRI for detection of pseudotumors and muscle atrophy, but it was superior for detection of joint effusion and tendinous pathologies. MARS MRI is more advantageous than USS for practical reasons, including preoperative planning and longitudinal comparison.


Link to article